| pH |
|
|
| Range |
-2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH |
| Resolution |
0.1 pH; 0.01 pH; 0.001 pH |
| Accuracy |
±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
| mV |
|
|
| Range |
±2000.0 mV |
| Resolution |
1 mV; 0.1 mV |
| Accuracy |
±0.2 mV ±1 last significant digit |
| Temperature |
|
| Range |
-20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.0 to 393.0 K |
| Resolution |
0.1 °C / 0.1 °F / 0.1 K |
| Accuracy |
±0.2 °C / ±0.4 °F / ±0.2 K |
| Relative mV offset range |
±2000.0 mV |
| pH Calibration |
|
| Calibration points |
Up to 5 |
| Type |
Automatic; Semiautomatic; Manual |
| Standard buffers |
Hanna and NIST – pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 |
| Custom buffers |
Up to 5 |
| Custom group |
Up to 5 |
| 1st calibration point |
Offset or Points (user setting) |
| Reminder |
Disabled; Daily: 0 min. to 23 hours and 59 min; Periodic: 1 min. to 500 days, 23 hours and 59 min. |
| Temperature Compensation |
Automatic or Manual |
| Reading |
|
| Modes |
Direct; Direct/Autohold |
| Stability criteria |
Accurate; Medium; Fast |
| Isopotential |
7.000 or 4.010 |
| Sampling rate |
1000 ms |
| pH View |
|
| Basic |
Measurement (pH, mV, Rel.mV, Abs.mV) Temperature, Stability status |
| Simple GLP |
Basic view information; Last calibration date, electrode offset, average slope, and electrode condition |
| Full GLP |
Simple GLP information and calibration point details Table Measurements updated every second are displayed in table Graph (Plot) pH (or mV) and temperature versus time graph can be panned or zoomed (pinch-to-zoom technology) |
| Logging |
|
| Type |
Automatic, Manual, Autohold |
| Number of records |
50,000 maximum per file; Stores at least 1,000 ,00 data points per user |
| Automatic interval |
1, 2, 5, 10, 30 seconds; 1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes |
| Sample ID |
Incremental mode |
| Export option |
.csv file format |
| Users |
Up to 9 users and admin. account (default) |
| Connectivity |
|
| USB-A |
2 ports for keyboard input or USB thumb drive |
| USB-C |
1 port for PC connectivity and USB-C type thumb drive |
| Wi-Fi & Ethernet |
FTP; Web server Log transfer and download; Email |
| RS232 |
Connecting peripherals |
| Power supply |
DC adapter 100-240AC to 24VDC 2.5A |
| Environment |
0 – 50 °C / 32 – 122 °F / 273 – 323 K maximum 95% RH non-condensing |
| Dimensions |
205 x 160 x 77 mm (8.0 x 6.2 x 3.0 “) |
| Weight |
Approximately 1.2 kg (26.5 lbs.) |