Shop

HI-6421 Advanced Dissolved Oxygen Optical Benchtop Meter

£1,781.50 (exc VAT) each

Download Product Page PDF
SKU: HI-6421-02 Categories: , , ,

The HI-6421 Advanced Benchtop Meter for Dissolved Oxygen is a laboratory-grade solution for the accurate measurement of DO, BOD, OUR and SOUR in both fresh and saltwater samples. This modern laboratory DO meter includes five different viewing modes, ethernet and WiFi connectivity and a comprehensive on-screen tutorial for ease-of-use.

HI-6421 – DO (HI-7641133 Optical probe)
 Range 0.00 to 50.00 mg/L (ppm) concentration; 0.0 to 500.0 % saturation
 Resolution 0.01 mg/L (ppm); 0.1 % saturation
 Accuracy from 0.00 to 20.00 mg/L (ppm) 1.5 % of reading or ±0.01 mg/L (ppm), whichever is greater
from 20.00 to 50.00 mg/L (ppm) ±5 % of reading
from 0.0 to 200.0 % saturation ±1.5 % of reading or ±0.1 %, whichever is greater
from 200.0 to 500.0 % saturation ±5 % of reading
HI-6421P – DO (HI-764833 Polarographic probe)
 Range 0.00 to 90.00 mg/L (ppm) concentration; 0.0 to 600.0 % saturation
 Resolution 0.01 mg/L (ppm); 0.1 % saturation
 Accuracy ±1.5 % of reading ±1, least significant digit
Barometric Pressure
 Range 450 to 850 mmHg; 600 to 1133 mBar; 60 to 133 kPa;
17 to 33 inHg; 8.7 to 16.4 psi; 0.592 to 1.118 atm
 Resolution 1 mmHg; 1 mBar; 1 kPa; 1 inHg; 0.1 psi; 0.001 atm
 Accuracy ±3 mmHg within ±15 % from the calibration point
±3 mmHg ±1 least significant digit
Temperature
 Range −20.0 to 120.0 °C
−4.0 to 248.0 °F
253.0 to 393.0 K
 Resolution 0.1 °C; 0.1 °F; 0.1 K
 Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K
DO Calibration
 Points DO optical One or two points automatic calibration at 100% (8.26 mg/L) and 0% (0 mg/L).
Single point manual using a value entered by the user in % saturation or mg/L.
DO polarographic Automatic-two points / User standard-single point
 Standards 0 and 100% saturation
 Reminder Disabled
Daily: 0 min. to 23 hours and 59 min.
Periodic: 1 min. to 500 days, 23 hours and 59 min.
Temperature Compensation Automatic or Manual
Salinity Compensation (Optical DO only) Automatic from 0 to 70 PSU (manually set)
0.0 to 70.0 % / 0.0 to 45.0 g/L / 0.0 to 42.0 psu
Reading
 Modes Direct
Direct/Autohold
 Stability criteria Accurate
Medium
Fast
 Isopotential 7.000 or 4.010
 Sampling rate 1000 ms
DO Views
 Basic Measurement (DO, Temperature)
Stability status
 Simple GLP Basic view information
Last calibration date, offset, average slope
 Full GLP Simple GLP information and calibration point details
 Table Measurements updated every second are displayed in table
 Graph (Plot) Measurement versus time graph can be panned or zoomed (pinch-to-zoom technology)
Logging
 Type Automatic, Manual, Autohold
 Number of records 50 000 maximum per file
Stores at least 1 000 000 data points per user
 Automatic interval 1, 2, 5, 10, 30 seconds
1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes
 Sample ID Incremental mode
 Export option .csv file format
Users Up to 9 users and the default administrator account
Connectivity
 USB-A 2 ports for keyboard input or USB thumb drive
 USB-C 1 port for PC connectivity and USB-C type thumb drive
 Wi-Fi & Ethernet FTP
Web server Log transfer and download
Email
 RS232 Connecting peripherals
Power supply DC adapter 100-240AC to 24VDC 2.5A
Environment 0 – 50 °C / 32 – 122 °F / 273 – 323 K maximum 95% RH non-condensing
Dimensions 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 “)
Weight Approximately 1.2 kg (26.5 lbs.)
Ordering Information HI-6421 is supplied with HI-7641133 optical dissolved oxygen probe (OPDO®); HI-764060 electrode holder; capillary pipette; 24 VDC power adapter; USB-C to USB-A cable; probe quality certificate; quick reference guide with instrument quality certificate.
HI-6421P is supplied with HI-764833 polarographic probe; HI-764060 electrode holder; capillary pipette; 24 VDC power adapter; USB-C to USB-A cable; probe quality certificate; quick reference guide with instrument quality certificate.